Sato Keiryoki Mfg. Co., Ltd.

61-9438-87 Heatstroke Wet Bulb Globe Temperature Meter with Calibration Report SK-150GT

Spec

  • Calibration point: Temperature 23℃ (isn't black sphere temperature), Humidity 60% (at 23℃)
  • Contents of calibration documents: Calibration report
  • Prior to measurement, Please set measurement environment.
  • Room measure (IN): Sun does not reach in a room or outdoor
  • WBGT index (℃) = 0.7 x humidity sphere temperature (℃) + 0.3 x black sphere temperature (℃)
  • Outdoor measurement (OUT): Outdoor where the sun reaches
  • WBGT index (℃) = 0.7 x humidity sphere temperature (℃) + 0.2 x black sphere temperature + 0.1 x dry sphere temperature (℃)] *in heatstroke under high temperature environment, Is the general term for failure contracting a disease collapsed balance of moisture and salt (sodium such as) within the body and heat regulation function to collapse. In the worst case, Variable might result in death, If you know the preventive methods it can be prevented.
  • With HOlD function
  • With high value (mAx)/minimum value (mIN) display function
  • With auto power OFF function
  • With screw hole can be fixed to camera tripod
  • Measurement environment setting function with WBGT index
  • WBGT index alarm: Settings range 20.0 - 49.9℃ the buzzer warning (release disabled)
  • Battery life: Continuous use about 600 hour (alarm operation none) *WBGT index: Acronym for WetBulbGlobeTempeRatuRe (wet sphere black sphere temperature), Index that is used to evaluate the thermal stress of workers (hot environment) defined in ISO7243/JISZ8504. Wet-bulb temperature, Black sphere temperature (radiant heat), Is one of the volume feeling temperature calculated from dry Sphere temperature. *WBGT index: Acronym for WetBulbGlobeTempeRatuRe (wet sphere black sphere temperature), Index that is used to evaluate the thermal stress of workers (hot environment) defined in ISO7243/JISZ8504. Wet-bulb temperature, Black sphere temperature (radiant heat), Is one of the volume feeling temperature calculated from dry Sphere temperature.
  • Measurement range: Temperature: 0.0 - 50.0℃, Black sphere temperature: 0.0 - 80.0℃, WBGT index: 0.0 - 50.0℃, Humidity: 10.0 - 95.0%Rh
  • Accuracy: Temperature (TA) ±0.6℃, Black sphere temperature (TG) ±2℃, WBGT index ±2℃, Humidity (RH) ±3%Rh (20 - 90% Rh at between 20℃ - 30℃) ±5%Rh (other than the left)
  • Body dimensions: (W) 71 x (H) 275 x (D) 30mm, Black sphere dimensions: φ60mm
  • Weight: Approximately 240g (including dry cell)

Order No. 61-9438-87
Model No. SK-150GT
JAN Code 0000000000000
Standard price JPY: 41,400 USD: 386.56
Excange rate 1USD= 107.10JPY
Valid price in Japan
Quantity 1set
Stock

in Japan

Qty.

Related Products